Workers in digital imaging and publishing use color management to achieve consistency throughout a workflow. The goal is to preserve the quality and accuracy of an image from capture to final reproduction. Each device in the workflow supports a different color space. The available color management systems profile the gamut capabilities of each device, and then limit the working color space to the gamut that is shared by all of them.
The workflow used by color metrologists is different. Many labs employ more
than one instrument to measure color or appearance and to assure quality control. Here we'll focus on measurements of spectral reflectance factors ("SRF") for uniformly colored samples. A spectrophotometer is used to measure SRF data throughout the visible spectrum, which extends from 400 - 700 nm.
Different instruments may not output the same SRF data based on measurements of the same transfer standard. In this situation tests of inter-instrument agreement may be used to measure the consistency between the participating instruments. Depending on the outcome of these tests, color management may be required to enable the meaningful sharing of SRF data.
Inter-instrument agreement cannot be achieved unless the test instruments employ the same geometry. See Color Geometry: A Matter of Degrees (another post within this blog) for a discussion of geometry.
In printing and the graphic arts metrologists who use sphere-based instruments employ the (d/8°) diffuse hemispherical specular-excluded ("SCE") geometry. Others prefer the (0°/45°) or (45°/0°) bi-directional geometry. Because gloss is important in the manufacture of paints and architectural ceramics, in those industries metrologists prefer the (t/8°) total hemispherical specular-included ("SCI") geometry, although certain segments of the coatings industry, namely coil coating, have standardized on 45°/0° geometry.
Among comparable instruments that use the same geometry, other factors may lead to differences in the measured SRF data. Before listing these factors, we stipulate that the instruments to be compared are maintained in good working order, and that suitable procedures are used to achieve consistent sample quality and to measure accurate SRF data.
Differences in the SRF values measured by comparable instruments may result from the following types of wavelength-dependent error: (1) Photometric scale. (2) Wavelength scale. (3) Bandwidth.
The standard multivariate regression procedure that is used to characterize these errors was described in the open literature more than 20 years ago. In a complementary approach, during 2009 the author developed transfer standards that are optimized for detecting these errors. For more information see Color Measurement Accuracy: Diagnostic Procedures (another post within this blog).
The multivariate regression procedure detects and measures the following errors: (1) Photometric zero (black point). (2) Photometric linear scale (100% reflectance line). (3) Photometric nonlinear scale. (4) Wavelength scale. (5) Bandwidth. This method cannot compensate for differences in instrument geometry, or for the errors that result from thermochromism or lateral diffusion. For a discussion of translucency see Color Measurement Accuracy: Translucent Materials (another post within this blog).
CyberChrome's OnColor Profiler® profiling system can analyze the errors listed above on most of the instruments that are in use today. This system includes software and 32 reduced-translucency transfer standards that are optimized for profiling. The OnColor Profiler® software compares the SRF data measured on each test instrument with data from a master instrument. The software generates correlation coefficients, or correction factors, that may be used to improve inter-instrument agreement with the master instrument. See How to improve Inter-instrument Agreement with Instrument Profiling (another post within this blog).
The author of this article is John W. Root Ph.D. of Mt. Baker Research who can be reached via email at jackroot@mtbakerresearch.com.
Posted by Elaine Becker on Thu, May 27, 2010
This blog post was authored by John W. Root of Mt. Baker Research.
During 2009 Mount Baker Research introduced transfer standards and procedures for
diagnosing instrument errors. Parts A and B of the Diagnostic Tile Set each contain 16 2 x 2 inch ceramic color standards. Part A supports instrument testing and the assessment of inter-instrument agreement in a multi-instrument environment. Part B enables licensed users of part A and CyberChrome's OnColor Profiler® software to upgrade to profiling.
The part A standards were selected to meet the following criteria: (1) Span the complete photometric range. (2) Span the widest possible high-chroma gamut in CIELAB color space. (3) Include only high quality standards with minimal translucency.
List of Supported Tests
1. Instrument Validation. For this test the part A standards must be pre-calibrated in the geometries of interest. (Three geometries are supported.) After the user measures spectral reflectance factors ("SRF") with the test instrument, statistical data analysis is used to diagnose its performance and compare it to a reference instrument.
2. Absolute Accuracy. To characterize the absolute spectrophotometric accuracy, the SRF data from the validation test may be compared with data measured at NRC for similar transfer standards. At present this test is supported for the (0°/45°) and (45°/0°) geometries.
3. Inter-laboratory Agreement. The agreement test is performed when the white calibration tile supplied with the test instrument is traceable to a standards laboratory other than NRC or NIST. This extension of the validation test measures differences in the photometric scales between the laboratories on which traceability is based. Based on the results of recent beta tests, this systematic error may significantly exceed the statistical uncertainty of the SRF measurements.
4. Instrument Revalidation. Repeating the validation test provides a basis for tracking instrument performance and for checking the user's methods for storage and maintenance of the transfer standards.
5. White Calibration Tile. If the white calibration tile supplied with the test instrument is translucent, significant measurement errors may result. Provided that LAV and VSAV sample ports are available, the part A set may be used to diagnose this problem.
6. Translucency Sensitivity. The part A set may be used to measure the sensitivity of the test instrument to SRF errors that result from lateral diffusion. This test requires the use of LAV and VSAV sample ports. For more details on translucency, refer to the blog "Color Measurement Accuracy: Translucent Materials"
7. OnColor Profiler®. Profiling improves inter-instrument agreement in a multi-instrument environment. This may be accomplished using the combined parts A and B standards together with CyberChrome's OnColor Profiler software. Part B of the tile set adds 16 more chromatic colors to better map the color gamut. For more information on the benefits profiling offers, read the white paper on this topic.
The author of this article is John W. Root Ph.D. of Mt. Baker Research who can be reached via email at jackroot@mtbakerresearch.com.